Not rated

Hygitech Helix Test

The Helix test checks the ability of the autoclave to sterilise hollow objects such as suction or irrigation tubes.

The tests should be carried out daily on a specific cycle, vacuum autoclave, so that the different sterilisation cycles of the day can then be carried out.

Tests are available in packs containing 100 or 250 strips + test capsule tube.

Packages:

1 x 100 pcs.

1 x 250 pcs.

Code: 1672.01
Brand: Hygitech
from 83,10 € up to –19 % from 66,90 € / packaging from 61,90 € excl. VAT
Category: Tests
helox2
from 83,10 € up to –19 %
Quantity: 100 pcs.
For order | 1672.01
Delivery to: 08/05/2024
83,10 €
66,90 € / packaging
61,90 € excl. VAT
Quantity: 250 pcs.
For order | 11128.01
Delivery to: 08/05/2024
133,10 €
110,90 € / packaging
102,70 € excl. VAT

Properties:

- The test consists of a 1.5 metre long tube open at one end. The other end is closed by a hermetically sealed capsule which contains a strip printed with an indicator ink.

- It is an essential test because dental practices have instruments with rather long cavities which, if not sterilised properly, can transmit bacteria and viruses to the patient and operator.

- The Helix test simulates the operation of the autoclave with regard to the sterilisation of porous loads and, in particular, is used to evaluate: the effectiveness of the pre-vacuum and thus the penetration of steam into the cavities; and the values of temperature and saturated steam pressure during the sterilisation phase.

- This test can only be used in Class B autoclaves.

Method of use:

- Subject this test to a sterilisation cycle of 134°C for 3 to 5 minutes to simulate the sterilisation performance of an autoclave for a "complex" medical device.

- The steam should penetrate the inside of the coiled tube, which is deliberately narrow, and cause a reaction on the strip printed with the indicator ink.

- If the ink on the strip turns completely black then the sterilisation is successful.

- However, if the ink is still yellow or partially yellow then the test is considered negative (this means that the vapour has not penetrated properly to the indicator).

- It is then advisable to call in a professional repairer who will be able to validate the origin of the anomaly (incorrect calibration, contact time too low, etc.).

Traceability:

To ensure traceability, it is important to keep each test result in a register.

The objective being to always prove the good functioning of the equipment in case of infection of a patient.

Be the first who will post an article to this item!

Only registered users may post articles. Please log in or register.

Do not fill out this field: